Installation
26
Changing the direction of display box
When the direction of the display box needs to be changed, change it by the following step.
Remove the fixing screws (M4
6 pcs) of the lid of the detector.
When the lid is fixed by the optical unit lock screws (with knobs) at 3 locations, remove the
screws.
Lift up the upper lid from the cylindrical part slowly by 15 cm with a handle.
Turn around the display box (It is possible to change it by the pitch by 30 degrees).
Close the upper lid slowly, and fix the screws (M4
6 pcs) securely with a torque wrench.
(Tightening torque:1.57 N•m)
Fig. 17 Changing the direction of display box
CAUTION
[TOP]
[Front]
(Before)
(After)
Hex socket head cap bolts
(M4
6 pcs)
Fixed screw of optical unit