● Appendix 1
pH calibration can be performed according to several buffer standards. The most common
standard is the US buffer standard. The default setup is US buffer standard. Alternative
standards that can be chosen are NIST, NIST2, DIN and CUST (Custom).
The pH buffers are temperature dependent i.e. the pH value changes with change in
temperature. The meter is intelligent to detect the temperature and pH value associated with
buffer when calibration is performed. It remembers all the temperature vs. pH value for all the
standards.
The pH vs. temperature values for the various standards are listed below:
< USA >
Temp. (°C)
pH 1.68
pH 4.01
pH 7.00
pH 10.01
pH 12.46
0
1.67
4.01
7.12
10.32
13.42
5
1.67
4.01
7.09
10.25
13.21
10
1.67
4.00
7.06
10.18
13.00
15
1.67
4.00
7.04
10.12
12.81
20
1.68
4.00
7.02
10.06
12.63
25
1.68
4.01
7.00
10.01
12.45
30
1.69
4.01
6.99
9.97
12.29
35
1.69
4.02
6.98
9.93
12.13
40
1.70
4.03
6.97
9.89
11.98
45
1.70
4.04
6.97
9.86
11.84
50
1.71
4.06
6.97
9.83
11.70
55
1.72
4.08
6.97
9.81
11.57
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......