![horiba LAQUA 2000 Series Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/horiba/laqua-2000-series/laqua-2000-series_instruction-manual_2162234062.webp)
Conductivity Setup
44
● P1.6 Calibration Clear Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
key,
P1.3 A.CAL
appears.
5. Press the
key,
P1.4 T.CFF
appears.
6. Press the
key,
P1.5 T.REF
appears.
7. Press the
key,
P1.6 C.CLR
appears.
8. Press the
ENT
key, default
NO
screen appears.
9. Use the
keys to change to
YES
. This erases the calibration data.
10. Press the
ENT
key,
CCLR DONE
screen appears briefly.
11.
P1.6 C.CLR
screen appears. This indicates completion of calibration clear setup.
key
key
key
key
key
key
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......