pH Setup
28
■
Setup
This section describes the setups in LAQUA 2000 series benchtop meters.
● P1 pH Setup
Using pH setup you can:
• Set pH buffer group
• Set pH resolution
• Set calibration alarm
• Erase calibration data
To set the pH function of PH2000, ION2000, or PC2000 meter, follow the procedure detailed
below:
Prerequisites
• Switch on the meter.
• Ensure that the meter is in pH measurement mode.
• Default pH buffer group in buffer setup is
USA
. You can change it to
NIST
,
NIST2
,
DIN
, or
CUST
(Custom)
if required.
• Erasing previous calibration data is recommended for accurate calibration. Default
calibration clear setup is
NO
but to erase the calibration data, change the setup to
YES
.
To return to the measurement mode, press the
key.
pH Buffer Group
pH Buffer Values (
25°C)
USA
1.68, 4.01, 7.00, 10.01, 12.45
NIST
1.68, 4.01, 6.86, 9.18, 12.45
NIST2
1.68, 4.01, 6.86, 10.01, 12.45
DIN
1.09, 3.06, 4.65, 6.79, 9.23, 12.75
CUST
Use up to 6 pH buffers that are 1.0 pH
apart for manual calibration
Note
Tip
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......