ORP Calibration
14
Calibration
1. After placing the ORP electrode in ORP standard solution, press the
key to switch
to mV mode.
2. Press the key.
3. Meter starts reading mV values.
4. Wait for the
to stabilize. When it stabilizes, there will be a sound indicating that
reading is stable.
5. Use the
and
keys to adjust the mV value to the ORP standard solution
value.
6. Press the
ENT
key to confirm and save calibration data.
7. Meter displays
CAL DONE
indicating end of the ORP/mV calibration procedure. mV
changes to RmV in measurement mode.
key
key
key
key