
Conductivity Setup
40
● P1.2 Conductivity Unit Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
ENT
key, default
UNIT S/cm
appears.
5. Use the
keys to change the conductivity unit to
S/m
.
6. Press the
ENT
key,
P1.2 UNIT
screen appears. This indicates completion of
conductivity unit setup.
key
key
key
key
key
key
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......