Salinity Calibration
23
Calibration
1. After placing the conductivity electrode in the salinity standard, press the key.
2. Meter starts measuring with a
blinking on screen.
3. Wait for the
to stabilize. When it stabilizes, there will be a sound indicating that reading is
stable.
4. Use the
and
keys to enter the salinity standard value.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
CAL
DONE
indicating end of the salinity calibration procedure. Electrode
icon lights up in measurement mode.
To view and/or print calibration data, press the
key when you are in the
CAL
mode. The
display scrolls through the calibrated salinity standard value and calibration factor.
key
key
key
key
Note