Conductivity Calibration
19
Auto Conductivity Calibration
1. After placing the conductivity electrode in the conductivity standard, press the
key.
2. Meter displays
AUTO CAL
and starts measuring various calibration values with a
blinking on screen.
3. Wait for the
to stabilize. When it stabilizes, there will be a sound indicating that
reading is stable.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
CAL
DONE
indicating end of the conductivity calibration procedure.
Electrode icon and calibrated conductivity standard icon light up in measurement mode.
6. Repeat for other calibration points as required.
7. You can calibrate one point for each range.
To view and/or print calibration data, press the
key when you are in the
CAL
mode. The
display scrolls through the calibrated conductivity standard values, calibration factors, and
average calibration factor.
key
key
Note