Ion Calibration
17
Calibration
1. After placing the ion electrode in standard solution, press the key.
2. Wait for
to stabilize. When it stabilizes, there will be a sound indicating that reading
is stable.
3. Press
and
keys to enter the standard solution concentration and adjust
the unit. Press the
key to adjust the decimal point location.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
CAL
DONE
that indicates end of the ion calibration procedure. Electrode
icon and number of calibration points done (e.g. 1 PT for 1 point, 2 PTS for 2 points)
light up in measurement mode.
6. Repeat above steps for other calibration points as required.
To view and/or print calibration data, press the
key when you are in the
CAL
mode.The
display scrolls through the calibrated values, segment slope(s) and average slope.
key
key
key
key
key
Note