Conductivity Calibration
20
Manual Conductivity Calibration
1. After placing the conductivity electrode in the conductivity standard, press the
key.
2. Meter displays
MANUAL CAL
and starts measuring with a blinking
on screen.
3. Wait for the
to stabilize. When it stabilizes, there will be a sound indicating that
reading is stable.
4. Use the
and
keys to enter the electrical conductivity value of the
conductivity standard. Press the
key to adjust the decimal point location.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
CAL
DONE
indicating end of the conductivity calibration procedure.
Electrode icon and number of calibration points done (e.g. 1 PT for 1 point, 2 PTS for 2
points) light up in measurement mode.
7. Repeat for other calibration points as required.
8. You can calibrate one point for each range.