Salinity Setup
48
● P3 SAL Setup
Using salinity setup, you can:
• Select salinity unit
• Select salinity type
• Erase calibration data
To set the salinity function of EC2000 or PC2000 meter, follow the procedure detailed below:
Prerequisites
• Switch on the meter.
• Ensure that the meter is in salinity measurement mode.
• Default salinity unit is
ppt
(parts per thousand). You can change the unit to
%
(percentage).
• Default salinity type is
NACL
(NaCl). You can change the salinity type to
SEA.W
(seawater).
• Erasing previous calibration data is recommended for accurate calibration. Default
calibration clear setup is
NO
but to erase the calibration data, change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip