Maintenance and Storage
68
● Cleaning the Conductivity Electrode
Always clean the electrode with deionized water after every measurement. When the response
is slow or residue from the sample adheres to the electrode, use the appropriate method below
to clean the electrode, and then clean again with deionized water.
● Conductivity Electrode Storage
If the electrode is stored in a dry state, the cell constant will change. Store with the black
electrode part immersed in deionized water, or with the protective cap filled with deionized
water and attached to the electrode.
To store the electrode for a long period, clean it well and attach the protective cap filled with
deionized water.
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Inorganic substance
Ethanol (keep the ethanol away from plastic parts)
Scale that formed
during long term
storage
A commercially available scale remover (neutral cleansing
solution for kitchen use, etc.) diluted by a factor of 100. If this
does not remove the scale, use diluted solution that contains
oxygen bleach (sodium percarbonate) or chlorine bleach
(sodium hypochlorite).
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......