Temperature Calibration
25
Calibration
1. After placing the electrode in a solution with known temperature, press the
key.
2. Press the
key to switch to temperature calibration mode. Meter displays measured
temperature value.
3. Use the
and
keys to adjust the temperature to the required value.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
CAL DONE
indicating end of the temperature calibration procedure.
key
key
key
key
Содержание LAQUA 2000 Series