8.0 SPECIFICATIONS
SPECIFICATIONS
HPB
HPA
Pressure Range
(2)
500-1200 hPa
0 - 17.6 psia
(Note: 1 hPa = 1 mBar)
Overpressure
(3)
4150 hPa
60 psia
Burst Pressure
(4)
48 x 10
3
hPa
700 psia
Operating Temperature Range
(5)
-40 to 85°C (-40 to 185°F)
Storage Temperature Range
-55 to 90°C (-67 to 194°F)
Media Compatibility
Suitable for non-condensing,
non-corrosive, non-combustable gasses
Accuracy over Operating Temperature Range
(1)
HPB100 and HPA100
± 0.8 hPa
± 0.06% FS
HPB200 and HPA200
± 0.4 hPa
± 0.03% FS
Stability
0.25 hPa per year
0.02% FS per year
Resolution
0.0011% FS
Temperature Readings
/- 1.0 C (at sensing element)
Outputs
TTL
(7)
(user configurable for either multi-drop or ring networking)
Power Requirements
6 to 26 VDC,
5-9mA depending upon configuration
Weight
5 oz. (142g)
Bus Addressing
Unique addresses 1 - 89, user assignable
Baud Rate
1200, 2400, 4800, 9600, 14400, 19200, 28800
Pressure Units
atm, bar, cmwc, ftwc, hPa, inHg, inwc, Kg/cm,
KPa, mBar, mmHg, MPa, mwc, psi user, lcom, pfs
Response Delay
Delay in ms = (1000/update rate) + 1ms
(6)
or 17ms, whichever is greater.
Sample Rate
8.33mS to 51.2 minutes per update
Mechanical Shock
1500 G for 0.5 msec half sine
(per MIL-STD-833D, M2002.3, Cond B)
Thermal Shock
24 cycles from -55 to +90°C (-67 to 194°F), 1 hr cycles
Vibration
0.5 in. or 20 Gs, 20 Hz to 2 KHz
(per MIL-STD-833D, M2007.2, Cond A)
Humidity Sensitivity
Negligible
Acceleration Sensitivity
Negligible
(1) Sum of worst case linerity, repeatability, hysteresis, thermal effects, and calibration errors from -40 to 85°C.
Consult Honeywell for other accuracy options.
(2) Consult Honeywell for other pressure ranges.
Note: HPB units meet accuracy specifications over 500 - 1200 hPa pressure range. HPB units are fully
characterized and will provide digitally compensated pressure values over the entire 0-1200 hPa range.
(3) Exposure to overpressure will not permanently affect calibration or accuracy of the unit.
(4) Exceeding burst pressure rating may result in media escape.
(5) Consult Honeywell for other temperature ranges.
(6) The user-selectable update rates very between 8.33 ms and 12 sec. Time to first reading is 150 ms.
(7) Unit default is multi-drop. Can be user-configured to ring networking.