13
Figure 2. Pressure Reading Control
Shaded blocks represent command codes executed in the microprocessor
• Single or Continuous
• ASCII or Binary
Pressure Values
P1, P2, P3, P4
Pressure
if c = 0 then no output
if c = 1 - 255 then add 'c' wait
cycles between inputs
IC = c
(IC=0 factory default)
Idle Count
if r = 0 then output every input
if r = 1 - 10 then skip r x 100
identical readings
RR = r
(RR=00 factory default)
Reading Rate
if d = A then output every input
if d = U then output only changes
if p = X then no watchdog timer
if p = W then use watchdog timer
For binary format mode:
if m = N then no checksum
if m = C then use checksum
if n = S then use standard format
if n = E then use extended format
OP
= d m n p
(OP=ANEX factory default)
Operating Mode
HPB Pressure Rate Conditioning
DS = dd Sn
(DS=00S0 factory default)
Dead Band and Sensitivity
User Compensation
Pressure
Input
TTL-Level
TD, RD
Digital
Outputs
I
= Rn or
I
= Mn
(
I
=M002 factory default)
range for n value: 1-120
Rn: n readings/sec
Mn: n x 100 msec/reading
Integration Time
if
I
= Rn
if
I
= Mn
A/D
Converter
Dead Band
DB = n x dd x 0.005% FS range
(dd= 0 - 60, n=1, 2, 4, or 8)
User Compensation
Press. = mX + b
(X=
m,
Z=
b)
X=nn, Z=nn
(X=1.000 factory default)
Delay
Output Control
Output Control
Output Mode
and Format
Integration
Display Units
Display Units
Scalar