0 or 1 provides a deadband multiplier to allow the user to specify a wider
set point null span where:
0 provides a 1x multiplier (deadband x 0.005% FS)
1 provides a 10x multiplier (deadband x 0.05% FS)
DU Display Units Control
Set the
D
isplay
U
nits for compensated pressure replies. A
USER
defined display unit can be set with this
command that causes the pressure output to equal the value input by the
U=
command multiplied by the
actual pressure in psi units. (Default:
DU=PSI
for HPA devices,
DU=HPA
for HPB devices)
Inquiry:
*dd
DU
EEPROM Store: *dd
WE
*dd
SP=ALL
Action:
*dd
WE
*dd
DU=
{
ATM
|
BAR
|
CMWC
|
FTWC
|
INHG
|
INWC
|
KGCM
|
KPA
|
MBAR
|
MMHG
|
MPA
|
MWC
|
PSI
|
USER
|
LCOM
|
PFS
}
Display Unit
PSI Multiplier
(1)
Units Description
ATM . . . . . . . . . . . . .
0.068046 . . . . . . . . . . . . . . . . . ambient atmosphere at sea level
BAR . . . . . . . . . . . . .
0.068948 . . . . . . . . . . . . . . . . . bar (105 newtons/meter2)
CMWC . . . . . . . . . . .
70.304 . . . . . . . . . . . . . . . . . . . centimeters of water column (4 °C)
FTWC . . . . . . . . . . . .
2.3065 . . . . . . . . . . . . . . . . . . . feet of water column (4 °C)
HPA . . . . . . . . . . . . .
68.948 . . . . . . . . . . . . . . . . . . . hectopascal (1 millibar)
INHG . . . . . . . . . . . .
2.0360 . . . . . . . . . . . . . . . . . . . inches of mercury (0 °C)
INWC . . . . . . . . . . . .
27.679 . . . . . . . . . . . . . . . . . . . inches of water column (4 °C)
KGCM . . . . . . . . . . .
0.070307 . . . . . . . . . . . . . . . . . kilograms/centimeter2
KPA . . . . . . . . . . . . .
6.8948 . . . . . . . . . . . . . . . . . . . kilopascal
MBAR . . . . . . . . . . .
68.948 . . . . . . . . . . . . . . . . . . . millibar (1 hectopascal)
MMHG . . . . . . . . . . .
51.714 . . . . . . . . . . . . . . . . . . . millimeters of mercury (0 °C) or torr
MPA . . . . . . . . . . . . .
0.0068948 . . . . . . . . . . . . . . . . megapascal
MWC . . . . . . . . . . . .
0.70304 . . . . . . . . . . . . . . . . . . meters of water column (4 °C)
PSI . . . . . . . . . . . . . .
1.0000 . . . . . . . . . . . . . . . . . . . pounds per square inch
USER . . . . . . . . . . . .
0.0010000-999.99 . . . . . . . . . . user supplied (see also
U=
command)
LCOM . . . . . . . . . . .
FS=60000 . . . . . . . . . . . . . . . . logic common value (2)
PFS. . . . . . . . . . . . . .
FS=100.000 . . . . . . . . . . . . . . . percent of full scale in 0.001% increments
Table 6. Display Units Options
(1)
From the CRC Handbook Of Tables For Applied Engineering Science
(2)
The
LCOM
display unit (logic common) automatically switches the digital pressure output to a scale which has 60000 as the
full scale output (±60000 for differential devices). The output will have a decimal point in the same position that it is in with the
original psi scale. For example, when a 17.6 psia device is switched to
DU=LCOM
, the full scale digital pressure output will be
60.000, instead of 17.600.