17
Figure 8. Idle Count (IC) Command Example
Figure 7b. Deadband and Sensitivity (DS) Command Examples
Figure 7a. Deadband and Sensitivity (DS) Command Examples
Pressure
(hPa)
1104
1106
1108
1102
1110
1112
1114
1116
1100
Setting DS=40 sets the deadband to
±40 x 0.005% FS = ±0.20%FS.
Then setting DS=C1 selects the center
deadband option with a 10x multiplier.
For a 1200 hPa full scale device, this
results in a ±2.4 hPa deadband window.
Deadband
(±2.4 hPa)
HPB pressure output
Actual pressure
Time (sec)
1.0
3.0
5.0
I = M5
DS=40C1
IC = 0
RR = 0 OP=A
Deadband - Center
Pressure
(hPa)
1104
1106
1108
1102
1110
1112
1114
1116
1100
Setting DS=40 sets the deadband to
±40 x 0.005% FS = ±0.20%FS.
Then setting DS=S1 selects the spread
deadband option with a 10x multiplier.
For a 1200 hPa full scale device, this results
in a ±2.4 hPa deadband window.
Deadband
HPB pressure output
Actual pressure
Time (sec)
1.0
3.0
5.0
I= M5
DS=40S1
IC = 0
RR = 0 OP=A
Deadband - Spread
Pressure
(hPa)
1110
1115
1120
1105
1125
1130
1135
1140
1100
Time (sec)
1.0
2.0
3.0
Idle Count
IC = 11 (skip 11)
IC = 4 (skip 4)
IC = 2 (skip 2)
IC = 0 (skip none)
1
2
3
4
5
6
7
8
9
10
11
12
Integration time
HPB pressure output
Actual pressure
I = M2
RR = 0
OP = A