3.0 COMMANDSQUICK REFERENCE
Example
Example
Input (1)
Response (2)
A=
Data String A
Write string A *00WE
(up to 8 characters, a <cr> indicates end-of-message.
*00A=2-8-95
More than 8 characters is an invalid write)
(Can store info. such as dates, readings, etc.)
Inquiry *00A=
?01A=2-8-95
BP
Baud Rate and Parity Setting
Set parameters *99WE
(no parity, 1200 baud)
*99BP=N1200
B=
Data String B
Write string B *00WE
(up to 8 characters, a <cr> indicates end-of-message,
*00B=123.4567
and more than 8 characters will not write to location)
(Can store info. such as dates, readings, etc.)
Inquiry *00B=
?01B=123.4567
CK
Check EEPROM
Inquiry *00CK
?01CK=OK
C=
Data String C
Write string C *00WE
(up to 8 characters, a <cr> indicates end-of-message,
*00C=This_is_
and more than 8 characters will not write to location
The C and D string can be used for a watchdog or reset
message(see MO command).
Inquiry *00C=
?01C=This_is_
DO Default Operating Parameters
Set DO parameters *00WE
(factory default setting)
*00DO=E
Inquiry *00DO
?01DO=E0NX
DSDeadband and S
ensitivity Control
Set DS parameters *00WE
set deadband to 20 x 0.005% = 0.10%FS )
*00DS=20
Inquiry *00DS
?01DS=00S0
DU
Display Units Control
Set DU parameters *00WE
(set units to in. Hg)
*00DU=INHG
Inquiry *00DU
?01DU=INHG
D=
Data String D
Write string D *00WE
(up to 8 characters, a <cr> indicates end-of-message,
*00D=A_HPB!!!
of more than 8 characters will not write to location)
The C and D string can be used for a watchdog or reset
message (see MO command).
Inquiry *00D=
?01D=A_HPB!!!
F=
Custom Full Scale Range
Set F= parameter *00WE
(set custom FS range to 10.5psi)
*00F=10.5
Inquiry *00F=
?01F=10.500
Note: See Section 5.10 for complete command descriptions.