C. Power Cord Replacement
Read carefully and utilize the check boxes to ensure all steps are performed. See figures below
for assistance.
Removal (with the console powered off)
□
1) Unplug console power cord from wall outlet.
□
2) Completely unwrap and remove power cord from cord management plate.
□
3) With #1 Philips screwdriver, loosen fastener on bottom of retaining bracket.
□
4) Disconnect power cord connector from power receptacle.
Installation (with the console powered off)
□
5) Firmly and completely insert power cord connector into power receptacle.
□
6) With #1 Philips screwdriver, tighten fastener on bottom of retaining bracket.
Verification of Installation
□
7) Plug power cord into wall outlet.
□
8) Turn console power on.
□
9) Verify console powers on, beeps and defaults to the biopsy mode.
□
10) Unplug power cord and wrap clockwise around cord management plate.
If you have questions or concerns with the removal, installation or verification of installation,
please contact the Suros Customer Support Department at 1-877-887-8767 for assistance.
Cord Management Plate
Power Cord
Power Receptacle
(Visible once the power cord
connector has been
removed)
Retaining Bracket
Power Cord Connector
Fastener