ATEC
Pearl, ATEC Sapphire and ATEC Emerald
Breast Biopsy and Excision System
LIT309:RevB
This document contains information that is confidential and proprietary to Suros Surgical Systems, Inc., a Hologic company. The document
and/or information contained herein may NOT be disclosed to any party without the express written consent of Suros. Covered by one or
more of the following patents and other patents pending: 6,758,824 and 6,638,235. © 2007 Suros Surgical Systems, Inc. ATEC, Suros and
Suros Compassionate Technologies are trademarks or registered trademarks of Suros in the United States and other countries. All rights
reserved.
28
Mount Replacement
Parts: From SR6418-01 Mount Kit
Tools:
Qty Part #
Description
5/32" Allen Wrench
4
SR6017-03
Mount, Isolation, Compressor
5/16” Allen Wrench
4
SR6017-04
Mount, Isolation, Vacuum Pump
11/32” Socket
8
SR6225-01
Washer, #8, Ext Tooth Lock, SS
5/16” Socket
8
SR6232-01
Nut, 8-32, Nylon Locknut, SS
1/4” Socket
8
SR6289-01
Nut, 6-32, Nylon Locknut, SS
3/16” Hex bit
8
SR6290-01
Washer, #6, Ext Tooth, SS
9/64” Hex bit
4
SR6291-01
Screw, 6-32, 5/8”, BH, SH, SS
11/32” Wrench
4
SR6292-01
Screw, 8-32 x 5/8”, BH, SH, SS
Loctite
5
SR6230-01
Cable Ties, 4”, Nylon
Side Cuts
1
SR6111-02
Muffler
Strap Wrench
Joint Compound
Vacuum Mount, top view. Vacuum Mount, bottom view.
Comp. Mount, top view. Comp. Mount, bottom view.
Removing the Compressor
1. Place a towel over the front edge of the front panel to protect the console.
2. Remove the 4 compressor screws with a 3/16” hex bit or Allen wrench.
3. Disconnect the brown wires from the capacitor.
4. Disconnect the black and white power wires (use needle nose pliers).
5. Disconnect the green grounding wire from the compressor with ¼” socket.
6. Snip the cable tie by holding the green grounding wire to the black and white
compressor power wires.
7. Remove the compressor assembly from the console.