A. Vacuum Line Assembly Replacement
Read carefully and utilize the check boxes to ensure all steps are performed. See figures below
for assistance.
Removal (with the console powered off)
□
1) Locate white lock collar where the vacuum line assembly attaches to the console.
□
2) Lift up on the vacuum line assembly approximately one-eighth inch, exposing a gap
between the white collar and black barb base.
□
3) Insert a small slotted screwdriver into the gap created, then press and hold down the
white collar.
□
4) While holding down the white collar, pull the vacuum line assembly up and out of the
vacuum line assembly connector.
Installation (with the console powered off)
□
5) Press the new vacuum line assembly into the vacuum line assembly connector.
□
6) Pull up to verify it has locked into the connector. The vacuum line assembly will have
small amount of vertical movement but should not pull free from connector.
Verification of Installation
□
7) Turn console power on.
□
8) An audible console beep will be heard; it will then default to the biopsy mode.
□
9) Plug the tip of the vacuum line assembly with thumb or finger.
□
10) Press the set-up button on the console.
□
11) Verify that the solid green vacuum ready light is illuminated, indicating that the system is
functioning properly.
If you have questions or concerns with the removal, installation or verification of installation,
please contact the Suros Customer Support Department at 1-877-88-SUROS (1-877-887-8767)
for assistance.
In-Line Hydrophobic Filter
Vacuum Line Assembly
Black Barb
Black Barb Base
White Lock Collar
Vacuum Line Connector