Power Switch Wiring
1. Press in the fingers on both sides of the switch and push the switch out of the top
panel slot.
a. Some switches have been glued into the top panel. Use any appropriate tools
that will help loosen and remove the switch.
2. Remove the wires from the power switch. Needle Nose Pliers may be helpful.
3. Discard old switch.
4. Attach the leads to the power switch as shown in the figure.
5. Route over the AC Filter.
6. Snap the switch into the panel.
a. Re-glue if glued originally.
7. Verify that the switch light operates.
3.6.19 5 Port Valve Replacement
Parts:
Tools:
Qty Part #
Description
#2 Phillips bit
2
SR6018-01
Valve, 5 Port
Torque Driver
White: From AC Filter
Black: From AC Filter
White: To 3 AC
Black: To 3 AC
Fingers
Lot #
5 Port Valves
5 Port Valve w/ Pilot
*Do not replace*