b. Insert 1 snubbing washer between the lock washer and the bottom of each
mount.
c. Insert 1 snubbing washer between the top of each mount and the compressor.
11. Feed the screw up from the bottom.
12. Torque to 60 in/lbs or until the mount starts to rotate with a 3/16” hex bit.
13. Install a cable tie attaching the compressor grounding wire and the compressor black
and white power wires together. Route the between the compressor and the pump.
14. Dress wires to prevent shorting or any type of damage.
3.6.17 Silencer Replacement
Parts: From Preventative Maintenance Kit Tools:
Qty Part #
Description
N/A
1
SR6325-01
Silencer
Silencer
1. Carefully unscrew the silencer by hand and remove the silencer from the brass
orifice. (May need to hold orifice)
2. Install the new orifice and tighten by hand.
3.6.18
Power Switch Replacement
Power Switch Replacement
Parts:
Tools:
Qty Part #
Description
Needle Nose Pliers
1
SR6071-01
AC Power Switch