Cenova Image Analytics Server 3.0 Installation & Service Manual
Chapter 3: Installing the System
Page 32
MAN-05205 Revision 003
3.4.2 Entering the Connectivity Settings
The System Settings – Connectivity screen lets you view networking and enter DICOM
information for the Cenova server. Any new entries or edits on this page will result in an
automatic restart of the DICOM and Processing services.
Note
If DICOM 6000 Interface and the Cenova application are running on the same server
make sure their respective local SCU and SCP AE Titles and ports are different.
To enter the server’s connectivity settings:
1.
Click
Connectivity
. The Connectivity screen appears.
Note
The entries for Host Name, IP Address, Subnet Mask, and Default Gateway are
derived from the values entered through the Windows Control Panel (see
Entering the Server Network Identifiers
on page 27).