Chapter 10: Appendix
199
Guidance and manufacturer’s declaration – electromagnetic immunity
The CARAT II pro ventilator
is intended for use in the electromagnetic environment spec-
ified below. The user of the
CARAT II pro ventilator
should ensure that it is used in such
an environment.
Immunity tests
IEC 60601
Test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic
discharge acc. to
(ESD) IEC 61000-
4-2
± 8 kV
contact
± 2, 4, 8, 15 kV
air
± 8 kV
contact
± 2, 4, 8, 15 kV
air
± 2, 4, 8, 15 kV
air display
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Radiated RF
disturbances acc.
to
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80% AM at 1kHz
10 V/m
Portable and mobile
communications
equipment (including
cables and antennas)
should be used no
closer to any part
of the CARAT II pro
home ventilator than
the recommended
separation distance of
0.3 m.
27 V/m
385 MHz PM: 18Hz
27 V/m
28 V/m
450 MHz
FM ± 5Hz: 1kHz sinus
28 V/m
9 V/m
710, 745, 780 MHz
PM:217Hz
9 V/m
28 V/m
810, 870, 930 MHz
PM: 18Hz
28 V/m
28 V/m
1720, 1845, 1970
MHz PM: 217Hz
28 V/m
28 V/m
2450 MHz PM:217Hz
28 V/m
9 V/m
5240, 5500, 5785
PM:217Hz
9 V//m
Electrical fast
transient/burst acc.
to IEC 61000-4-4
± 2 kV for
power supply lines
± 1 kV for
input/output lines
± 2 kV for
power supply lines
± 1 kV for
input/output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...