INSTRUCTIONS BEFORE COMMISSIONING
Only a properly functioning device may be used on the patient.
After connection to the mains, the device performs a hardware
test (see on page 42). Make sure that you hear two diffe-
rent alarms and the LEDs are flashing. If you do not hear the
alarm sounds or if not all LEDs are flashing, then the device
must not be used.
•
Locate the device in such a way that the mains plug is easily
accessible and can be unplugged quickly in the event of a
potential hazard.
•
Do not use the device if the housing or the cable of the
device or the power supply are damaged.
USING OF OXYGEN
•
Please be sure to observe the user’s manual of the manu-
facturer or distributor from whom you obtain the oxygen.
•
If the patient is supplied with oxygen via the device, the
FiO
2
should be measured.
•
CARAT II pro offers a FiO
2
measurement via the optional
FiO
2
sensor. We recommend using this particular sensor
exclusively.
•
The FiO
2
sensor contains a caustic liquid. Avoid skin or eye
contact if there is a sensor leak! Replace the sensor.
•
The oxygen supplied must not exceed a pressure of 1000 hPa
and a flow of 15 l/min. The oxygen must be dosed using
an external flow meter.
•
When supplying oxygen, please ensure that only dry gas
(FiO
2
) is used. Increased residual moisture may lead to
device defects. If necessary, a humidifier can be connected
between the air outlet of the device and the patient.
•
The connection between the FiO
2
connection and external
FiO
2
source must be absolutely airtight. Otherwise, leakage
losses may occur during ventilation.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...