142
Chapter 6: Device operation
STATISTICS
The statistics screen contains statistical evaluations of the following ventilation
parameters.
• Minute volume
• Frequency
• SpO
2
• Leak Rate
• Tidal Volume
• I:E Ratio
The evaluation of the ventilation parameters is based on percentiles. Percentiles
are the dispersion measurement of the statistical data distribution during ventila-
tion sessions.
An example based on frequency
25
20
15
10
0
08
f
[ bpm ]
during 95 % of the day's therapy duration the frequency was
≤
24 bpm
and 5 % was
>
24 bpm
during 75 % of the day's therapy duration the frequency was
≤
22 bpm
and 25 % was
>
22 bpm
during 50 % of the day's therapy duration the frequency was
<
15 bpm
and during 50 %
>
15 bpm
during 25 % of the day's therapy duration the frequency was
≤
12 bpm
and 75 % was
>
12 bpm
during 5 % of the day's therapy duration the frequency was
≤
10 bpm
and 95 % was
>
10 bpm