98
Chapter 5: Ventilation modes
FLOW RAMP
VCV
AVCV
V-SIMV
The flow ramp can be set in the volume controlled modes as a constant inspi
-
ration flow (level 1
), as decelerating inspiration flow (level 2
, level 3
,
level 4
) or as an accelerating/decelerating inspiration flow (level 5
). The flow
curve can be changed depending condition of lung (compliance and restistance).
INSPIRATION TRIGGER
APCV
PSV
PSV-S
AVCV
P-SIMV
V-SIMV
The inspiration trigger is a volume trigger. It specifies the inspiration efforts of
patient equired to obtain pressure or volume support from the ventilator in the
case of spontaneous breathing.
Please always consider the patient’s clinical record when setting the trigger
levels in order to avoid the risk of auto triggering.
Once the set inspiratory trigger threshold (see Table 5) is reached, the device starts
the inspiratory phase.
Table 5:
Thresholds
inspiratory
trigger
Level
Volume threshold value
Flow change in 30 ms
1
5 ml (± 20 %)
1,2 l/min
2
10 ml (± 20 %)
1,5 l/min
3
15 ml (± 20 %)
1,8 l/min
4
20 ml (± 20 %)
2,1 l/min
5
25 ml (± 20 %)
2,4 l/min
6
30 ml (± 20 %)
2,7 l/min
7
35 ml (± 20 %)
3,0 l/min
8
40 ml (± 20 %)
3,3 l/min
9
45 ml (± 20 %)
3,6 l/min
10
50 ml (± 20 %)
3,9 l/min
To avoid auto triggering, which are cause by artifacts of the single line patient cir-
cuit expiration valve, the "SMART Function" can be set (see page 100).
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...