174
Chapter 9: Routine checks and maintenance work
PERFORM THE BATTERY QUICK TEST
The battery quick test verifies whether the device in case of a power failure can
be operated on battery for at least one hour. When the device is used the battery
quick test must be performed out by the provider or user every 6 months. How-
ever, a monthly check is recommended.
1. Make sure that the batteries are fully charged (100%).
2. Disconnect the device from the mains and operate the device for 1 hour on
battery power.
3. The test is positive if after 1 hour the battery capacity is > 10 % and the alarm
"Low Internal Battery" has not sounded. If the battery capacity has fallen below
10 % and the alarm "Low Internal Battery" has sounded, the batteries must
be replaced by an authorized service technician.
4. Charge the batteries up to 100 % by operating the devices via mains.
REPLACE THE BATTERIES
The batteries must be replaced every 2 years by an authorized service technician.
The procedure is described in the CARAT pro service manual.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...