Chapter 5: Ventilation modes
85
Important settings
Tips
• Alarm limit
"High Inspiration Volume"
• Alarm limit
"Low Inspiration Volume"
• Alarm limit
"High Expiration Volume"
• Alarm limit
"Low Expiration Volume"
• Alarm limit "High Frequency"
• Adjustable "Trigger Lock" of the
inspiration trigger
• Minimum (Ti Min) and maximum
(Ti Max) inspiration times are adjust-
able
0
5
10
15
20
25
20
10
0
IPAP
PS
PEEP
t (s)
P
10 s
∆t
10 s
∆t + t
exp
Ⓟ
Ⓜ
Ⓜ
t (s)
V
Ⓐ
IPAP
= 20 hPa
PS
= 12 hPa
PEEP
= 8 hPa
Frequency
= 6 bpm
Spontaneous breathing
Expected time window = 5 s
Ⓜ
Mandatory ventilation
Ⓟ
Pressure supported ventilation
Ⓐ
Pressure controlled assisted ventilation
Figure 38:
P-SIMV mode diagram