102
Chapter 5: Ventilation modes
MINIMUM VOLUME
PCV
APCV
PSV
PSV-S
The minimum volume is the minimum tidal volume, which is considered as a vol-
ume guarantee during a pressure controlled ventilation.
ADDITIONAL PRESSURE
PCV
APCV
PSV
PSV-S
The additional pressure is graduately added by 2 hPa to the IPAP or PS pressure
until the set additional pressure is reached, if the minimum volume has not been
reached in order to guarantee the set minimum volume.
SIGH FUNCTION
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
If using a volume controlled mode and the sigh function is activated, 150% of the
set volume is administered in the inspiration phase at each 100th breath. If using
a pressure controlled mode, the IPAP is increased so that approx. 150 % of the
messured tidal volume of the previously breath is reached. This may help to reo-
pen or keep open any collapsing pulmonary alveolar.
If in the P-SIMV and V-SIMV modes the 100th breath is a spontaneous pressure
supported breath, only the next ventilation will be performed as a "sigh".
MAX. PRESSURE
VCV
AVCV
V-SIMV
The "Max. Pressure" setting limits the pressure towards the top and does affect
the "Tidal Volume" setting.
MIN. PRESSURE
VCV
AVCV
V-SIMV
The "Min. Pressure" setting limits the bottom pressure.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...