Chapter 10: Appendix
183
Storage and transport conditions
Relative humidity
5 % ... 95 %, non-condensing
Air pressure range
250 hPa ... 1100 hPa
Storage conditions
Store in a dry, vibration-free place, in an upright
position; store device and accessories in their origi-
nal packaging.
Sound pressure range of audible alarm signal (at 1 m distance)
Lowest value
55 dBA, Level 1
Medium value
60 dBA, Level 2
Highest value
65 dBA, Level 3
Resistance in the first event of a fault
Inspiratory pressure of the
device at the patient con-
nection port at 60 l/min
Single line patient circuit
Double line patient circuit
2.9 hPa
3.7 hPa
Expiratory pressure of the
device at the patient con-
nection port at 60 l/min
Single line patient circuit
Double line patient circuit
2.1 hPa
5.0 hPa
Measurement conditions
Device with single line
patient circuit (art. no.
00014967) without any
accessories
Device with double line
patient circuit (art. no.
00007969) without any
accessories
Application parts
Mask, Tracheal cannula, Endotracheal tube, SpO
2
finger clip sensor
Technical requirements for accessories
Oxygen inlet
Connection type
Quick-connect coupling
Pressure
≤
1000 hPa
Flow
≤
15 l/min
Bacterial filter
Connections
22 / 15 mm cone (according to EN1281-1)
Resistance
< 2.3 hPa at 60 l/min
Compressible volume
< 66 ml
Internal volume
< 200 ml
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...