Chapter 6: Device operation
113
EXPLANATION OF TOOLBAR ICONS
Icon
Meaning
Clinic mode active
Home mode active
Alarm active
Red icon
→
High priority alarm
Yellow icon
→
Medium priority alarm
Turquoise icon
→
Low priority alarm
Audio alarm paused
The audible alarm has been paused for 2 min. The audible alarm of even
a new alarm event will also be paused for 2 min. The audible alarm may
be deactivated by pressing the alarm key before an alarm event occurs.
Pressing the key again reactivates the audible alarm in case an alarm
event has occurred.
1:45
Counter "Audio alarm paused counter"
Indicates how much longer the audible alarm will be paused.
Key lock activated
The functions of all controls are disabled, except for the ON/OFF and
alarm key.
A FiO
2
sensor is connected and "FiO
2
-Monitoring" setting has been set
to "Internal" in the system screen.
The "FiO
2
-Monitoring" setting has been set to "External" in the system
screen. Measurement of oxygen concentration is performed using an
external FiO
2
-Monitoring unit.
A FiO
2
sensor is connected, the "FiO
2
-Monitoring" setting has been set
to "Internal" in the system screen and oxygen is supplied.
The "FiO
2
-Monitoring" setting has been set to "External" in the system
screen. Measurement of oxygen concentration is performed using an
external FiO
2
-Monitoring unit and oxygen is supplied.
Spontaneous breathing detected
The device has detected spontaneous breathing by the patient. This
triggered the inspiration trigger. The icon will remain visible during inspi-
ration and will shut off with the beginning of the expiration.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...