Chapter 2: Safety warnings
25
•
We recommend the use of the tube systems tested and
approved for use by the manufacturer. Using other tube
systems may lead to aberrant results.
•
Manufacturer tested and approved accessories are recom-
mended for the device. If other accessories are used, this
may lead to insufficient ventilation or the use of hazardous
materials may lead to further, secondary complications.
•
Equipment which are not part of the ventilation system
must be not be connected.
•
When a nasal or full face mask is used for non-invasive
ventilation, this non-vented mask must not contain any
expiration opening.
•
If used with a single line patient circuit, the controlled expira-
tion valve must not meet any resistance during exhalation and
must allow quick ventilation of the ventilation tube system.
•
In order to ensure patient safety, the device must be oper-
ated in such a way that all adjustable alarms are activated
and adjusted to the patient.
•
Alarms must not be ignored. They indicate conditions that
require an immediate action.
•
After changing any ventilation or alarm parameters, the
tubing system, or the system configuration (e.g., humidifier,
filter, oxygen addition, etc.), the effectiveness of ventilation
and alarms should be checked.
•
Every two years a safety-related test and maintenance is
required for the ventilator.
•
In case of exessive agitation on the part of the patient,
there is a risk of hyperventilation in all ventilation modes
with inspiration triggering.
•
The device must not be steam-sterilized in an autoclave.
•
Filters and other parts that are connected to the device
must be regularly replaced. Please dispose of the used
parts according to the regulations for used medical material
and/or the local environmental protection rules.
•
The connection of accessories or other components to the
respiratory system of the ventilator can lead to increased
expiratory pressure at the patient connection opening.
Содержание CARAT II pro
Страница 2: ......
Страница 22: ...22 Chapter 1 Introduction ...
Страница 40: ...40 Chapter 3 Description of device ...
Страница 106: ...106 Chapter 5 Ventilation modes ...
Страница 148: ...148 Chapter 6 Device operation ...
Страница 168: ...168 Chapter 8 Cleaning and disinfection ...
Страница 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...
Страница 194: ...194 Chapter 10 Appendix Name Article number Figure Brief instructions for CARAT II pro 50000645 Carrying case 00004875 ...