72
Chapter 4: Commissioning
100
80
60
40
20
0
3
5
7
9
11
13
15
FiO
2
[%]
Flow O
2
[l/min]
f = 25 bpm
MV = 2.5 l/min
MV = 5.0 l/min
MV = 7.5 l/min
MV = 10.0 l/min
Example
100
80
60
40
20
0
3
5
7
9
11
13
15
FiO
2
[%]
Flow O
2
[l/min]
f = 25 bpm
78
11
11 l/min
MV = 2.5 l/min
MV = 5.0 l/min
MV = 7.5 l/min
MV = 10.0 l/min
At an oxygen flow rate of 11 l/min and a minute ventilation of approximately
6 l/min, an oxygen concentration of approximately 78 % can be reached.