29
Procedure for Switching the Channel
Switching the shield
When the connection method is selected, the shield wire is connected to the specified connection
destination.
Specified shield switching setting (shielding connection destination)
Module
Connection method
Shield connection destination
SW9001
2-wire
TERMINAL 1
, LOW terminal
4-wire
GND
SW9002
2-wire
TERMINAL 1
, LOW terminal
4-terminal pair
TERMINAL 3
, sense shield terminal
The destination for shielding connection can be switched as necessary (for each slot).
Any time the connection method for a slot is changed, the switching shield will be reset to the
specified value. To set the switching shield to a value other than the specified value, be sure to set
the switching shield after setting the connection method.
Destination for shield wire connection to be selected
Module
Connection destination terminal
SW9001
Not connected
GND
TERMINAL 1
, LOW terminal
TERMINAL 2
,
TERMINAL 3
, source LOW terminal
TERMINAL 3
, sense shield terminal
TERMINAL 1
, LOW te
TERMINAL 3
, sense shield terminal
SW9002
Not connected
GND
TERMINAL 1
, LOW terminal
TERMINAL 3
, sense shield terminal
For the SW9001, the [
TERMINAL 1
, LOW te
TERMINAL 3
, sense shield terminal] shielding
connection destination terminal should only be used when connecting a BT3562, BT3562A series
(BT3562, BT3563, BT3561A, BT3562A, BT3563A) and DM7275 series (DM7275, DM7276) at the
same time when the DM7275/DM7276 contact check function is not operating properly.
Once the connection destination is specified, it is retained until the connection method is changed.
You do not have to specify the connection destination every time the channel is switched.
When the switching shield is set, all channels are set to open.
The source terminal for
TERMINAL 2
and sense terminal are common for
TERMINAL 3
.
Item
Communication command
Setting
procedure
:SYSTem:MODule:SHIeld <
slot No.
>,<OFF/GND/TERMinal1/TERMinal2/
TERMinal3/T1T3>
Setting
example
Set the shield wire connection destination for
SLOT 1
to GND.
:SYST:MOD:SHI 1,GND
Set the shield wire connection destination for
SLOT 5
to “not connected”.
:SYST:MOD:SHI 5,OFF
3
Channel Switching
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...