Installing the Module
2.2 Installing the Module
“Before installing the module” (p. 5)
The following settings are initialized when you install a new module into a slot or change the type of
module installed in a given slot.
• Connection method
• Shield switching
• Channel delay
• Scan list
Required items: Phillips screwdriver (No. 2), antistatic gloves
Front
1
1
Touch the
GND
terminal with
bare hands.
2
Wear antistatic gloves.
Rear
3
3
Turn off the device.
Front
4
Remove all the measurement
cables connected to the device
and module.
(To prevent electric shock and short
circuit of the measuring object.)
Blank panel
Screw
(M3 × 6 mm)
5
Loosen the two screws
(M3 × 6 mm) and then remove
the blank panel.
Store the blank panel and screws.
You need the screws when using the
device after removing the module.
2
Preparation for Measurements
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...