![Hioki SW9001 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/hioki/sw9001/sw9001_instruction-manual_2137434037.webp)
33
Measurement between Two Instruments
Control example using communication command
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE4
Set the
SLOT 1
connection method to 4-wire (connect to
TERMINAL 2
).
[SW1001] :CLOSE 101
Select
SLOT 1
, CH1.
[SW1001] *OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
*
OPC?
query.
[BT3562A] :READ?
Execute single measurement using BT3562A.
[BT3562A] 1.0258E-3
Receive measured values.
[SW1001] :CLOSE 102
Select the next CH2.
...
(Repeat until CH8.)
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE2
Set the
SLOT 1
connection method to 2-wire (connect to
TERMINAL 1
).
[SW1001] :CLOSE 112
Select
SLOT1
, CH12 (4-wire CH1 sense).
[SW1001] *OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
*
OPC?
query.
[DM7276] :READ?
Execute single measurement using DM7276.
[DM7276] +03.00
Receive measured values.
[SW1001] :CLOSE 113
Select the next CH13 (4-wire CH2 sense).
...
(Repeat until CH19.)
Example: Measure the internal resistance of 8 batteries using BT3562A and measure the OCV
using DM7276. Then measure the external potential for batteries using the DM7276.
(Use
SLOT 1
, 4-wire, CH1 to CH8 for internal resistance and OCV measurement.
Use
SLOT 2
, 2-wire, CH1 to CH8 for external potential measurement for batteries.)
1
Connect BT3562A to
TERMINAL 2
and DM7276 to
TERMINAL 1
.
Instrument
Measurement item
Module Connection
method
Channel selection Terminal to be used
BT3562A
Internal resistance
SW9001
4-wire
SLOT 1
, CH1
TERMINAL 2
SLOT 1
, CH2
SLOT 1
, CH3
⁞
SLOT 1
, CH8
DM7276
OCV
SW9001
2-wire
SLOT 1
, CH12
TERMINAL 1
SLOT 1
, CH13
SLOT 1
, CH14
⁞
SLOT 1
, CH19
DM7276
External potential
measurement for
batteries
SW9001
2-wire
SLOT 2
, CH1
TERMINAL 1
SLOT 2
, CH2
SLOT 2
, CH3
⁞
SLOT 2
, CH8
2
Set the connection method of
SLOT 1
to 4-wire.
The signal is connected to
TERMINAL 2
.
3
Channel Switching
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...