![Hioki SW9001 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/hioki/sw9001/sw9001_instruction-manual_2137434058.webp)
54
Timing Chart
(2) Scan operation reset
When the SCAN_RESET signal is input during scan operation, the scan operation is stopped and
the channel relay is opened. When the scan signal is input with the scan operation reset, the scan
operation is restarted from the beginning.
SCAN
OFF
SCAN_RESET
OFF
OFF
CLOSE
ON
ON
OFF
ON
T6
Scan channel
Channel n
Channel n+1
Channel n+2
Open
Scan stop
Open
Open
Item
Description
Time
T1
SCAN signal pulse width
1 ms or longer
T2
Relay settling time (when it is opened)
SW9001: 5 ms, SW9002: 5 ms
T3
Relay settling time (when it is closed)
SW9001: 5 ms, SW9002: 5 ms
T4
Channel delay time (user setting)
0 ms to 9999 ms (based on the setting)
T5
CLOSE signal pulse width
1 ms to 100 ms
As per the
:IO:PULSe:TIME
command setting
T6
SCAN_RESET signal pulse width
1 ms or longer
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...