103
Message Reference
(8) RS-232C settings
Set and query RS-232C communication speed
Syntax
Command
:SYSTem:COMMunicate:RS232C:SPEED <9600/19200/38400>
Query
:SYSTem:COMMunicate:RS232C:SPEED?
Response
<9600/19200/38400>
Description
Sets the RS-232C communication speed. The speed setting is immediately changed.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the speed
setting is fixed to 9600 bps regardless of the settings specified here.
The query returns the set value regardless of the communication setting mode switch
state.
Other RS-232C communication parameters are fixed to data bits: 8; stop bit: 1; parity
check: none; flow control: none.
Note
An execution error occurs during scan operation (command).
(9) LAN settings
Set and query communication command port No.
Syntax
Command
:SYSTem:COMMunicate:LAN:CONTrol <1
-
9999>
Query
:SYSTem:COMMunicate:LAN:CONTrol?
Response
<1 to 9999 (NR1)>
Description
Specifies the TCP/IP port where communication commands are received via LAN.
The setting is enabled after
:SYSTem:COMMunicate:LAN:UPDate
(or restart) is executed.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the port No.
is fixed to 23 regardless of the settings specified here.
The query returns the set value regardless of the communication setting mode switch state.
Example
SYST:COMM:LAN:CONT 23
SYST:COMM:LAN:UPD
SYST:COMM:LAN:CONT?
23
Note
An execution error occurs during scan operation (command).
Set and query default gateway
Syntax
Command
:SYSTem:COMMunicate:LAN:GATeway <Value 1>,<Value
2>,<Value 3>,<Value 4>
Query
:SYSTem:COMMunicate:LAN:GATeway?
Response
<Value 1>,<Value 2>,<Value 3>,<Value 4>
<Value 1> = 0 to 255 (NR1)
<Value 2> = 0 to 255 (NR1)
<Value 3> = 0 to 255 (NR1)
<Value 4> = 0 to 255 (NR1)
<Value 1> to <Value 4> indicate the address.
Description
Sets the default gateway address.
The setting is enabled after
:SYSTem:COMMunicate:LAN:UPDate
(or restart) is executed.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the address
is fixed to 0.0.0.0 (none) regardless of the settings specified here.
The query returns the set value regardless of the communication setting mode switch state.
Example
SYST:COMM:LAN:GAT 192,168,0,100
SYST:COMM:LAN:UPD
SYST:COMM:LAN:GAT?
192,168,0,100
Note
An execution error occurs during scan operation (command).
8
Communication Function
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...