96
Message Reference
(2) Connection method
Set and query connection method
Syntax
Command
:SYSTem:MODule:WIRE:MODE <
Slot No.>,<Connection method>
Query
:SYSTem:MODule:WIRE:MODE?
Slot No.
>
Response
<Connection method>
<Slot No.> = 1 - 3 (NR1) (SW1001)
1 - 12 (NR1) (SW1002)
<Connection method> = WIRE2/WIRE4/TP4
WIRE2: 2-wire
WIRE4: 4-wire
TP4: 4-terminal
Description
Sets the connection method for each slot. Connection method selection determines the
connection terminal for the instrument to which each channel is connected.
The connection method settings vary depending on the module type.
When the connection method is set, all relay is opened.
Module Parameter that
can be set
Connection
method
Number of
channels
Instrument
example
Model
Terminal
Connection
cable
SW9001
WIRE2
2-wire
22
Voltmeter
DM7276
TERMINAL 1
L4930
WIRE4
4-wire
11
Battery tester
BT3562A
TERMINAL 2
L2108
Resistance Meter RM3545
SW9002
WIRE2
2-wire
6
Voltmeter
DM7276
TERMINAL 1
L4930
TP4
4-terminal
pair
6
Battery tester
BT4560
TERMINAL 3
L2004
LCR meter
IM3590
Example
:SYST:MOD:WIRE:MODE 1,WIRE2
Set the connection method for
SLOT 1
to 2-wire (use
TERMINAL 1
for measurement).
:SYST:MOD:WIRE:MODE 12,TP4
Set the connection method for
SLOT 12
to 4-terminal pair (use
TERMINAL 3
for
measurement).
:SYST:MOD:WIRE:MODE? 5
WIRE4
The connection method for
SLOT 5
is 4-wire.
Note
•
When the connection method is selected, the shield switching is set to the specified
value. Set the shield switching as necessary.
•
Specifying a slot without a module will result in an execution error.
•
An execution error occurs during scan operation (command).
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...