4.8 Changing the Measurement Method (Auto/Manual)
69
3
Cha
pter 4
Se
ttings
4
All measurement items cannot be turned off.
At least one item must be on.
(An attempt to turn off all items, will cause one item to go on.)
Shown when network A, C, D, E, F or G is selected.
Line phase in leakage current measurement between
enclosure and line
Conducts measurement with voltage of the L
(live) side of power supplied to [LINE IN].
Conducts measurement with voltage of the N
(neutral) side of power supplied to [LINE IN].
EX.:Automatic measurement item setting
measurement between enclosure and line
Shown when network B2 is selected
These two settings cannot be set to OFF at the same time.
Measurement with 110% voltage application
(Other test condition)
Conducts measurement with voltage that is in
the same phase as power supplied to [LINE IN].
Conducts measurement with voltage that is in
the reversed phase of power supplied to [LINE
IN].
Conducts measurement turning off power supply
connected to [LINE IN].
*
*
*
EX.:Automatic measurement item setting
screen for network B2 and patient leakage
current (SIP/SOP) measurement
Shown when network B2 is selected
These two settings cannot be set to OFF at the same time.
*The "N" key indicates positive phase and "R" negative
phase.
Measurement with 110% voltage application
(Special test condition)
Conducts measurement with voltage that is in
the same phase as power supplied to [LINE IN].
Conducts measurement with voltage that is in
the reversed phase of power supplied to [LINE
IN].
*
*
EX.:Automatic measurement item setting
screen for network B2 and patient leakage
current (Metal accessible part) measurement
At least 1 item will be on.
It is not possible to turn off all items.
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......