1.2 Features
8
To test electrical equipment for leakage current, it is necessary to use a measurement network that simu-
lates the human body in compliance with applicable standards or laws.
The instrument includes the 8 measurement networks below.
The use of these networks enables instrument compliance with other standards.
Once a measurement network is selected, the instrument displays a leakage current measurement mode
corresponding to the applicable standard or law.
This function makes it possible to continue testing without shutting down the equipment under test when
polarities are changed, which shortens the duration of the test.
(Be sure to use an isolation transformer when using power on polarity switching.)
The instrument provides a rated current of 20 A and a rated voltage of 250 V.
1.2 Features
Wide range of measurement networks in compliance with standards
and laws
1. Network A
:For compliance with Electrical
Appliance and MaterialSafety Law
2. Network B1 :For JIS T0601-1:1999
3. Network B2 :IEC 60601-1:2005 3rd
Complies with JIS T 0601-1:2012
4. Network C
:For IEC60990
5. Network D
:For UL
6. Network E
:For general-purpose 1
7. Network F
:For general-purpose 2
8. Network G
:For IEC 61010-1
Leakage current measurement mode
1. Earth leakage current
2. Touch current (Enclosure - Earth)
3. Touch current (Enclosure - Enclosure)
4. Touch current (Enclosure - Line)
5. Patient auxiliary current
6. Patient leakage current (Patient connection - Earth)
7. Patient leakage current (external voltage on a SIP/SOP
*
)
8. Patient leakage current (external voltage on a specific F-type applied part)
9. Patient leakage current
(external voltage on metal accessible part not protectively earthed)
10. Total patient leakage current (Patient connection - Earth)
11. Total patient leakage current (external voltage on a SIP/SOP
*
)
12. Total patient leakage current (external voltage on a specific F-type applied part)
13. Total patient leakage current
(external voltage on metal accessible part not protectively earthed)
14. Free current (Enclosure - Enclosure)
15. Enclosure - Earth leakage current
16. Enclosure - Enclosure leakage current
17. Enclosure - Line leakage current
18. Patient leakage current I
19. Patient leakage current II
20. Patient leakage current III
*
SIP(Signal Input Part)/SOP(Signal Output Part)
Power on polarity switching function
Rated current 20 A
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......