4.6 Setting the Allowable Value
61
3
Cha
pter 4
Se
ttings
4
Set the allowable value (upper-limit leakage current value) for the measurement value and maximum
value.
According to the status (normal condition, single-fault condition) of the equipment under test, two values
can be set.
The allowable value setting is based on "Numeric value x Factor."
Use the allowable value setting to provide a margin or compensate for measurement error.
The factor is set to "100%" by default. (Input range 1 to 100% )
Set a lower limit value when required.
• Setting a lower limit value for all measurement modes.
See
"7.9 Setting Lower Limit Values (factor)" (p. 130)
• Setting a lower limit value for specific measurement modes.
See
"Turning lower limit values on and off for specific leakage current measurements" (p. 63)
The instrument indicates the result (PASS/FAIL/LOW indication) of comparison between the measure-
ment value and allowable value.
In automatic measurement, the judgement result is output from EXT I/O.
See
"9.2 Connecting to the EXT I/O Terminal" (p. 247)
The buzzer that sounds based on the judgement result can be disabled.
See
"7.13 Setting the Beep Sound" (p. 135)
4.6 Setting the Allowable Value
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......