
1.6 Types of Leakage Current Measurement
21
1
Chapter 1
Overview
*
Measurement with a disconnected protective earth conductor is applicable only to Class-I equipment.
List of Leakage-Current Measurement Items
Measurement
Normal
condition
Single-
fault
condition
(Description of fault)
(Other conditions)
Touch current
Enclosure - Line
1. One wire in the power
cord is disconnected.
2. protective earth conduc-
tor is disconnected
*
Situation other than touch current
(Enclosure - Line) is applicable
1. Functional grounding wire is disconnected
2. Grounding wire for patient connection and power
supply circuit for measurement are disconnect-
ed
3. A voltage that is 110% of the rated voltage is ap-
plied between an isolated signal input/output
section and earth (Not medical equipment)
Touch current
•
Enclosure - Earth
•
Enclosure - Enclosure
Earth leakage current
1. One wire in the power
cord is disconnected.
1. Functional grounding wire is disconnected
(Class I only)
2. Grounding wire for patient connection and power
supply circuit for measurement are disconnect-
ed (Class I only)
Patient leakage cur-
rent (Patient connec-
tion - Earth)/
(Patient leakage
current I)
1. One wire in the power
cord is disconnected.
2. The protective earth
conductor is discon-
nected.
1 Functional grounding wire is disconnected
2. Grounding wire for patient connection and power
supply circuit for measurement are disconnect-
ed
Patient leakage
current
(external voltage on a
SIP/SOP)/
(Patient leakage
current II)
1. One wire in the power
cord is disconnected.
2. The protective earth
conductor is discon-
nected.
1. Functional grounding wire is disconnected
2. Metal accessible part not protectively earthed
and grounding wire is disconnected
3. A voltage that is 110% of the rated voltage is ap-
plied between an isolated signal input/output
section and earth
Patient leakage cur-
rent (external voltage
on a specific F-type
applied part)/
(Patient leakage
current III)
1. A voltage that is 110% of the rated voltage is ap-
plied between an F-applied part and earth.
(Does not qualify as a single fault condition un-
der IEC 60601-1: 2005 3rd Edition.)
2. Metal accessible part not protectively earthed
and grounding wire is disconnected
3. Functional grounding wire is disconnected
Patient leakage cur-
rent (external voltage
on metal accessible
part not protectively
earthed)
1. The protective earth
conductor is discon-
nected.
1. Applied to metal accessible part not protectively
earthed
2. Functional grounding wire is disconnected
Patient auxiliary
current
1. One wire in the power
cord is disconnected.
2. The protective earth
conductor is discon-
nected.
Functional grounding wire is disconnected
G
e
ne
ra
l e
lec
tric
al
eq
u
ipm
e
n
t
Med
ica
l electrical eq
uip
m
en
t
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......