4.8 Changing the Measurement Method (Auto/Manual)
68
The selected network, class and measurement mode determine what measurement items will appear.
See
Appendix2 "List of instrument status, other test condition and special test condition"(p.A4)
Setting automatic measurement items
1.
On the automatic measurement setting screen, press
to display the automatic measure-
ment item setting screen.
2.
Turn automatic measurement for each measurement item ON or OFF.
Each time a key is pressed, the indication switches between ON and OFF.
Equipment with an internal power supply cannot be set.
See
"Line phase in leakage current measurement
between enclosure and line" (p. 69)
1. Power supply polarity
Indicates "positive phase."
Indicates "negative phase."
2.Condition of equipment under test
(power supply)
Sets "Normal condition."
Sets a "Single fault condition"
(power supply lead is disconnected).
Sets a "Single fault condition"(protective
earth terminal is disconnected).
Automatic measurement item setting screen
1
2
See
"Measurement with 110% voltage application (Other
test condition)" (p. 69)
Other test condition (Other applied voltages)*
2
Sets application of 110% of voltage,
positive phase.
Sets application of 110% of voltage,
negative phase.
Sets application of 110% of voltage, OFF.
*
1
*
1
Other conditions
Applies a voltage to the L line.
Applies a voltage to the N line.
Sets application of 110% of voltage,
positive phase.
(When selecting network B1.)
Sets application of 110% of voltage,
negative phase.
(When selecting network B1.)
*
1
*
1
Measurement current can be selected when automatic
measurement item is selected only when network B1 or B2
is selected and patient leakage current I mode, patient leak-
age current (Patient connection - Earth) or patient auxiliary
current is also selected. (In any other mode, AC + DC is
automatically selected and cannot be changed)
*
1
The "N" key indicates positive phase and "R" negative
phase.
*
2
When selecting network B2.
Target current
Sets AC measurement.
Sets DC measurement.
ON : Sets automatic
measurement item.
OFF : Does not set
automatic measurement item.
See
"Measurement with 110% voltage application (Special
test condition)" (p. 69)
Special test condition
(Specific applied voltage)*
2
Sets application of 110% of voltage,
positive phase.
Sets application of 110% of voltage,
negative phase.
*
1
*
1
Return to
setting
screen.
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......