4.4 Setting the Measurement Range (Auto/Hold)
58
About Peak Over "!"
•
When measuring in Hold
(fixed) range setting, if a por-
tion of measuring current goes
over the input allowable range
for the circuit, "!" together with
"
FAIL
" may be displayed on
the measuring screen. "!"
means that some of the instan-
taneous values go over the
input allowable range. In this
case, please select a lower
sensitivity range to measure.
• When measuring in Auto
(automatic) range, if some of the instantaneous values goes over the input
allowable range and then a lower sensitivity range is chosen, the judgement
result may be out of the guaranteed accuracy range.
• "!" will be displayed only in ammeter mode and not in voltmeter mode. "!" infor-
mation is not available for RS-232C or GP-IB communication or printers. In case
of displaying "!", "
OVER FLOW
" and "
Range Up
" will also be displayed.
• An error screen like the one
shown in the figure appears
when an excessive current input
is made to the instrument.
Check the equipment under test
and the wiring and press the
"Ret" key.
You must press the
to
start operation.
!
Содержание ST5540
Страница 2: ......
Страница 6: ...Contents iv Index i ...
Страница 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Страница 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Страница 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Страница 144: ...7 14 Setting the Backlight 138 ...
Страница 284: ...10 14 Measurement Networks 278 ...
Страница 290: ...11 4 Instrument Disposal 284 ...
Страница 320: ...Appendix4 External Dimensions A30 ...
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......