1.6 Types of Leakage Current Measurement
16
This is electric current that flows through an "
applied part
human body
ground" path.
Measure medical instruments with a
signal input/output section
and
applied parts
other than F-type. Mea-
sure assuming that medical equipment with a broken
signal input/output section
is connected. Instead of
connecting damaged medical equipment to generate a
single-fault condition
, input a voltage that is 110%
of the rated voltage.
(Use the T3 terminal on the instrument)
IEC 60601-1:2005 3rd Edition stipulates that the connection of damaged medical equipment does not
constitute a
single-fault condition
, but handles it as a
normal condition
.
Patient leakage current measurement (external voltage on a SIP/SOP)
JIS T0601-1:1999 treats patient leakage current II as patient leakage current I plus a
single fault condition
.