22
Setting the Amplitude
4.3 Setting the Amplitude
1
2
1
Select the
[Amplitude]
setting.
2
Press a function key (
F1
to
F5
) or select a button with the mouse.
Sweep
p.41
Up-Down
p.18
Tenkey entry
p.19
Valid amplitude setting range: 0 V p-p to 20 V p-p (in 1 mV p-p increments)
The output voltage (amp offset) range for which accuracy is guaranteed for each unit is
as follows:
U8793: -10 V to +15 V
MR8790: -10 V to +10 V
If the value obtained by adding the amplitude and offset exceeds the range within which accuracy
is guaranteed, part of the output waveform will be clamped as follows:
U8793: Upper limit of +16 V, lower limit of -11 V
MR8790: Upper limit of +14 V, lower limit of -14 V
www.
.com
1.800.561.8187