![Hioki MR8790 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/hioki/mr8790/mr8790_instruction-manual_2137099017.webp)
14
Connecting Wires to the External Control Terminals (U8793)
2.4 Connecting Wires to the External Control
Terminals (U8793)
The following describes how to connect wires to the unit’s external control terminals.
Procedure
You will need
Single-strand wire
Stranded wire
10 mm
Compatible wires: Single-strand wire
I
0.65 mm (22 AWG)
Twisted wire 0.32 mm
2
(22 AWG)
Usable wire:
Single-strand wire
I
0.32 to
I
0.65 mm (28 to 22 AWG)
Stranded wire 0.08 to 0.32 mm
2
(28 to 22 AWG)
Strand
diameter
I
0.12 mm (per strand)
Standard stripped wire length: 9 to 10 mm
Tool for manipulating terminal buttons: Flat-head screwdriver (with a shaft
diameter of
I
3 mm and a tip width of 2.6 mm)
Connection method
1
2
3
1
Depress the terminal button with a
ÀDWKHDGVFUHZGULYHURURWKHUWRRO
2
Insert the wire into the wire connection
hole while continuing to depress the
button.
3
Release the button.
The wire will be locked in place.
Terminal
no.
Operation
1
OUT: Outputs a LOW-level signal during
waveform output.
2
IN: When using the program function,
inputting a LOW-level signal from an
external device will cause the unit to cancel
the hold state and transition to the next
step.
3
GND terminal
(same potential as the Memory HiCorder’s
GND)
www.
.com
1.800.561.8187